HD2717T Transmitter, indicator, ON/OFF regulator, temperature and humidity data logger with interchangeable probe, Merk : DeltaOhm[23 Jul. 2009, 1:32:06]
Keterangan
Description
The instruments of the HD2717T... series are transmitters, indicators, and ON/OFF
regulators with data logging functions, they measure temperature and humidity.
The main feature of these instruments is their interchangeable probe. The probe
can be replaced by the user without process interruption. Thus, the probe can be
calibrated or repaired at a later time.
The instrument is available in three different versions: with horizontal probe (S.TO),
vertical probe (S.TV) or with remote probe (S.TC), having the probe connected to the
electronics by means of a cable of various lengths. The S.TO and S.TV probes are
made of stainless steel AISI304, the S.TC probes can be of stainless steel AISI304
or POCAN (plastic material).
The probe is factory calibrated and ready to use, it is provided with a SICRAM2
module which stores the calibration data of the probe, allowing the interchangeability
of the probes.
The instruments measure:
• Temperature in Celsius or Fahrenheit temperature scale
• Relative humidity
and calculate:
• Absolute humidity
• Mixing Ratio
• Dew point
All models have both current and voltage outputs.
Some models are fitted with two control relays and one alarm relay, configurable by
the user.
All models are fitted with a multistandard RS232/RS485 serial port and an auxiliary
RS232C standard serial output. The RS485 serial output allows the management of
more than one device in a network.
The models HD2717T... can be with or without LCD. The display shows on the first
line the relative humidity or a derived parameter and on the second line the temperature
in degrees Celsius or Fahrenheit.
The data logger function allows to store the measures with a selectable storage
interval.
The instrument setup remains permanently stored, while the real time clock is protected
by an apposite Lithium battery against temporary mains voltage interruptions.
The power supply can be chosen, at the time of placing the order, between 24Vac/
dc or universal 90...240Vac.
Instrument versions and available probes
Display
HD2717Tx-0x Absent
HD2717Tx-Dx Custom LCD
Relay
HD2717Tx-x0 Absent
HD2717Tx-xR 2 control relays with change-over contact.
1 alarm relay with normally open contact.
Type of probe
HD2717T.xx Instrument with vertical probe S.TV or probe with cable S.TC.
HD2717TO.xx Instrument with horizontal probe S.TO.
Probes complete with SICRAM2 module for instruments HD2717T.xx
S.TV
The material of the S.TC...probes can be chosen between stainless steel AISI304
or POCAN plastic material.
S.TC1.2 Probe L=130mm with cable 2m
S.TC1.2P Probe L=130mm with cable 2m (POCAN probe)
S.TC1.5 Probe L=130mm with cable 5m
S.TC1.5P Probe L=130mm with cable 5m (POCAN probe)
S.TC1.10 Probe L=130mm with cable 10m
S.TC1.10P Probe L=130mm with cable 10m (POCAN probe)
S.TC2.2 Probe L=330mm with cable 2m
S.TC2.2P Probe L=330mm with cable 2m ( POCAN probe)
S.TC2.5 Probe L=330mm with cable 5m
S.TC2.5P Probe L=330mm with cable 5m (POCAN probe)
S.TC2.10 Probe L=330mm with cable 10m
S.TC2.10P Probe L=330mm with cable 10m (POCAN probe)
Probes complete with SICRAM2 module for instruments HD2717TO.xx
S.TO1 horizontal probe L= 130mm
S.TO2 horizontal probe L= 330mm
Technical specifics (@ 24Vac and 20°C)
Inputs
Temperature Sensor Pt100 classe 1/3 DIN
Working range of the
sensor
-50 ... +200°C (-58 ...+392°F)
Humidity Relative humidity %RH 5 ... 98%RH
Working range of the
sensor in temperature
-50 ... +150°C
(Special configurations up to 180°C
available on request)
Dew point TD -50 ... +100°C
Absolute humidity 0 ... 600g/m3
Mixing ratio 0 ... 2000g/kg of dry air
Accuracy of the
measured physical
quantities
Temperature Pt100 ±0.25°C
Relative humidity %RH
±2.5%RH (5...90%RH)
±3.0%RH (90...98%RH)
Accuracy of the calculated
physical quantities
See table in the following
chapter
Response time
3min with grid protection (at 20°C
and 0.5m/s)
Outputs
Communications Type RS232C and RS485 Multidrop
Baud Rate
9600 baud
57600 baud non-permanent
Physical quantities Measured Temperature, relative humidity
Calculated
Dew point, absolute humidity,
(mixing ratio).
Analog
outputs
Output types
4...20mA; 0...20mA
0...10Vdc; 2…10Vdc
Load resistance
Current output: 500© max
Voltage output: 100k© min
Resolution 16bit
Accuracy analog outputs ±0.05% f.s. @20°C
In case of measuring error
(exceeding of the operating
limits, faulty or not connected
probe,....)
Idc = 22mA
Vdc = 11V
Relay Working relay
2 x 3A/250Vac Load resistance,
1 change-over contact
Alarm relay
1 x 3A/250Vac Load resistance,
1 with normally open contact
Instrument
Power supply Versions
24Vdc / 24Vac 50…60Hz, ±10%
90 ... 240Vac 50…60Hz
Average consumption 3W
Data logger Storage capacity 9000 samples in max. 256 sessions
Storage type Circular memory
Stored parameters
Temperature, relative humidity, dew point,
absolute humidity, mixing ratio, analog outputs
1 and 2, relay status1, 2, 3.
Storage interval
1, 2, 5, 10, 20, 60 seconds, 2 and 4 minutes
Real time clock Type
Real time with
Lithium buffer battery
Accuracy ±1min/month
Software
DeltaLog12
for Windows® 98 to XP operating systems
Display LCD Custom segment LCD
Ambient working
conditions of the
electronics
Operating temperature -20…+60°C
Relative humidity 0…90%RH - No condensate
Static working pressure
of the sensors
12 bar max.
Storage temperature -30…+80°C
Housing L x H x W 143x154x61
Weight 600g
Material ABS
Degree of protection Electronics IP65
Accuracy of the calculated physical quantities
The accuracy of the calculated physical quantities depends on the accuracy of the
relative humidity and temperature calibration. The provided values refer to an accuracy
of ±2.5%RH, ±0.25°C, 1013.25mbar.
Models with vertical probe (S.TV) or separated probe with cable (S.TC)
HD2717T.00: Model of without display and without relay.
HD2717T.0R: Model of without display, with configurable control relays (2) and alarm relay (1).
HD2717T.D0: Model of with custom display, without relay.
HD2717T.DR: Model of with custom display, with configurable control relays (2) and
alarm relay (1).
Models for horizontal duct probe (S.TO)
HD2717TO.00: Model without display and without relay.
HD2717TO.0R: Model without display, with configurable control relays (2) and alarm relay (1).
HD2717TO.D0: Model with custom display, without relay.
HD2717TO.DR: Model with custom display, with configurable control relays (2) and
alarm relay (1).
Interchangeable temperature and humidity probes with SICRAM2 module, vertical
S.TV or with cable S.TC
S.TV: Vertical probe. Length of stem 130mm.
The material of the S.TC...probes can be chosen between stainless steel AISI304 or
POCAN plastic material.
S.TC1.2: Probe with cable. Length of stem 130mm, length of the cable 2m.
S.TC1.2P: Probe with cable. Length of stem 130mm, length of the cable 2m. Made of POCAN.
S.TC1.5: Probe with cable. Length of stem 130mm, length of the cable 5m.
S.TC1.5P: Probe with cable. Length of stem 130mm, length of the cable 5m. Made of POCAN.
S.TC1.10: Probe with cable. Length of stem 130mm, length of the cable 10m.
S.TC1.10P: Probe with cable. Length of stem 130mm, length of the cable 10m. Made
of POCAN.
S.TC2.2: Probe with cable. Length of stem 330mm, length of the cable 2m.
S.TC2.2P: Probe with cable. Length of stem 330mm, length of the cable 2m. Made of POCAN.
S.TC2.5: Probe with cable. Length of stem 330mm, length of the cable 5m.
S.TC2.5P: Probe with cable. Length of stem 330mm, length of the cable 5m. Made of POCAN.
S.TC2.10: Probe with cable. Length of stem 330mm, length of the cable 10m.
S.TC2.10P: Probe with cable. Length of stem 330mm, length of the cable 10m. Made
of POCAN.
Interchangeable temperature and humidity probe with SICRAM2 module, horizontal
S.TO
S.TO1: Horizontal probe for instrument HD2717TO.xx. Length of stem 130mm.
S.TO2: Horizontal probe for instrument HD2717TO.xx. Length of stem 330mm.
Accessories
RS27: RS232 null-modem serial connection cable with 9 poles sub-D 9 female connector
and 3 pole connector for COM AUX port. (Included in the supply of the
instruments without display).
DeltaLog12: Further unit of software for PC connection, data download, instrument
setup, and management of an instrument network. For operative systems Windows
® 98 to XP.
HD75: 75%RH saturated solution for checking the relative humidity sensor, complete