RTD THERMOMETERS HD2107.1 AND HD2107.2 SENSORS: Pt100, Pt1000, Ni1000, NTC
Keterangan
The HD2107.1 and HD2107.2 are portable instruments with a large LCD display. They measure the temperature using Pt100 or Pt1000 immersion, penetration, contact or air probes. The sensor can be a Pt100 3 or 4 wires, Pt1000, Ni1000 or NTC 2 wiresThe probes are fi tted with an automatic detection module, with the factory calibration settings already being memorized inside.The HD2107.2 instrument is a datalogger. It memorizes up to 80,000 samples which can be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0. The storing interval, printing, and baud rate can be confi gured using the menu.The HD2107.1 and HD2107.2 models are fi tted with an RS232C serial port and can transfer the acquired measurements in real time to a PC or to a portable printer.The Max, Min and Avg function calculate the maximum, minimum or average values.Other functions include: the relative measurement REL, the HOLD function, and the automatic turning off that can also be disabled.The instruments have IP67 protection degree.INSTRUMENT TECHNICAL CHARACTERISTICSInstrument Dimensions (Length x Width x Height) 185x90x40mm Weight 470g (complete with batteries) Materials ABS, rubber Display 2x41D2 digits plus symbols Visible area: 52x42mm
Operating conditions Operating temperature -5…50°C Warehouse temperature -25…65°C Working relative humidity 0…90%RH without condensation Protection degree IP67Power Batteries 4 1.5V type AA batteries Autonomy 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20¼A Mains Output mains adapter 9Vdc / 250mAMeasuring unit °C - °F - °K
Security of stored data Unlimited, independent of battery charge conditions Time Date and time Schedule in real time Accuracy 1min/month max departure Measured values storage - model HD2107.2 Type 2000 pages containing 40 samples each Quantity Total of 80000 samples Storage interval 1s…3600s (1hour)Serial interface RS232C Type RS232C electrically isolated Baud rate Can be set from 1200 to 38400 baud Data bit 8 Parity None Stop bit 1 Flow Control Xon/Xoff Serial cable length Max 15m Immediate print interval 1s…3600s (1hour)USB interface - model HD2107.2 Type 1.1 - 2.0 electrically isolatedConnections Input module for the probes 8-pole male DIN45326 connector Serial interface and USB 8-pole MiniDin connector Mains adapter 2-pole connector (positive at centre)
Measurement of temperature by Instrument Pt100 measurement range -200…+650°C Pt1000 measurement range -200…+650°C Ni1000 measurement range -50…+250°C NTC measurement range -30…+120°C Resolution 0.01°C in the range ±199.99 °C 0.1 °C in the remaining fi eld Accuracy ±0.01°C Drift after 1 year 0.1°C/year
TECHNICAL DATA OF PROBES AND MODULES EQUIPPED WITH INSTRUMENTTemperature probes Pt100 sensor with SICRAM moduleModelTypeApplication rangeAccuracyTP472IImmersion-196°C…+500°C±0.25°C (-196°C…+350°C) ±0.4°C (+350°C…+500°C) TP472I.0Immersion-50°C…+400°C±0.25°C (-50°C…+350°C) ±0.4°C (+350°C…+400°C) TP473P.0Penetration-50°C…+400°C±0.25°C (-50°C…+350°C) ±0.4°C (+350°C…+400°C) TP474C.0Contact-50°C…+400°C±0.3°C (-50°C…+350°C) ±0.4°C (+350°C…+400°C) TP475A.0Air-50°C…+250°C±0.3°C (-50°C…+250°C) TP472I.5Immersion-50°C…+400°C±0.3°C (-50°C…+350°C) ±0.4°C (+350°C…+400°C) TP472I.10Immersion-50°C…+400°C±0.3°C (-50°C…+350°C) ±0.4°C (+350°C…+400°C) TP49AImmersion-70°C…+400°C±0.25°C (-50°C…+350°C) ±0.4°C (+350°C…+400°C) TP49ACContact-70°C…+400°C±0.25°C (-50°C…+350°C) ±0.4°C (+350°C…+400°C) TP49APPenetration-70°C…+400°C±0.25°C (-50°C…+350°C) ±0.4°C (+350°C…+400°C) TP875Globe thermometerØ 150mm-10°C…+100°C±0.25°C
Common characteristicsResolution 0.01°C in the range ± 199,99°C 0.1°C in the remaining fi eldTemperature drift @ 20°C 0.003%/°C
4 wire Pt100 and 2 wire Pt1000 ProbesModelTypeApplication rangeAccuracyTP47.100Pt100 4 wires-50…+400°CClass ATP47.1000Pt1000 2 wires-50…+400°CClass A
Common characteristicsResolution 0.01°C in the range ± 199.99°C 0.1°C in the remaining fi eldTemperature drift @ 20°C Pt100 0.003%/°CPt1000 0.005%/°CORDER CODESHD2107.1K: The kit is composed of the instrument HD2107.1, connection cable for serial output HD2110CSNM, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. The probes must be ordered separately.HD2107.2K: The kit is composed of the HD2107.2 datalogger, connection cable HD2101/USB, 4 1.5V alkaline batteries, operating manual, case and DeltaLog9 software. The probes must be ordered separately.HD2110CSNM: 8-pole connection cable MiniDin - Sub D 9-pole female for RS232C.HD2101/USB: Connection cable USB 2.0 connector type A - 8-pole MiniDin.DeltaLog9: Software for download and management of the data on PC using Windows 98 to XP operating systems.AF209.60: Stabilized power supply at 230Vac/9Vdc-300mA mains voltage.S’print-BT: On request, portable, serial input, 24 column thermal printer, 58mm paper width.
Probes complete with SICRAM moduleTP472I: Immersion probe, sensor Pt100. Stem Ø 3 mm, length 300 mm. Cable length 2 metres.TP472I.0: Immersion probe, sensor Pt100. Stem Ø 3 mm, length 230 mm. Cable length 2 metres.TP473P.0: Penetration probe, sensor Pt100. Stem Ø 4mm, length 150 mm. Cable length 2 metres.TP474C.0: Contact probe, sensor Pt100. Stem Ø 4mm, length 230mm, contact surface Ø 5mm. Cable length 2 metres.TP475A.0: Air probe, sensor Pt100. Stem Ø 4mm, length 230mm. Cable length 2 metres.TP472I.5: Immersion probe, sensor Pt100. Stem Ø 6mm, length 500 mm. Cable length 2 metres.TP472I.10: Immersion probe, sensor Pt100. Stem Ø 6mm, length 1,000mm. Cable length 2 metres.TP49A: Immersion probe, sensor Pt100. Stem Ø 2.7mm, length 150mm. Cable length 2 metres. Aluminium handle.TP49AC: Contact probe, sensor Pt100. Stem Ø 4 mm, length 150mm. Cable length 2 metres. Aluminium handle.TP49AP: Penetration probe, sensor Pt100. Stem Ø 2.7mm, length 150mm. Cable length 2 metres. Aluminium handle.TP875: Globe thermometer Ø 150 mm with handle, complete with SICRAM module. Cable length 2 metres.Temperature probes without SICRAM moduleTP47.100: Direct 4 wires Pt100 sensor immersion probe. Probe’s stem Ø 3mm, length 230mm. Connection cable 4 wires with connector, length 2 metres.TP47.1000: Pt1000 sensor immersion probe. Probe’s stem Ø 3mm, length 230mm. Connection cable 2 wires with connector, length 2 metres.TP47: Only connector for probe connection: Pt100 direct 3 and 4 wires, Pt1000 2 wires and Ni1000 2 wires