Vapor Tracer[28 Nov. 2008, 22:44:55]
Keterangan
Detector Type:Ion Trap Mobility Spectrometer (ITMS)
Analysis Time:Default 8 seconds
Sample Acquisition:Air collection for vapor, surface wipe for trace particles
Warmup Time:20 minutes, approximately
Input Power (external power supply):110 to 120 VAC or 200 to 240 VAC (auto ranging), 47 to 63 Hz, 3.2A (max)
Output (external power supply):Maximum 130W (15VDC @ 8.6A)
Power Dissipation:37.5W (typical), 65W (during warmup period)
Input Power:(VT2 DC power connector) 15 Volt DC @ 2.5A
Replacement Fuse:6.3A, 250V Type T
Signal Processing:Variable integration time
Plasmagram component peak deconvolution
Recognition on multiple peaks, multiple controlled drugs and explosives
Output to bar graph display or time-of-flight plasmagram display
Storage Temperature:-4 to 140ºF (-20 to 60ºC)
Operating Temperature:Explosives Operations: 14 to 131ºF (-10 to 55ºC)
Narcotics Operations: 14 to 122ºF (-10 to 50ºC)
Operating Humidity:0-95% non-condensing
Optional Laptop Computer:Full specification provided at time of sale
Power Supply:115/230 VAC, 50/60 Hz
2 VDC
1.5-hour battery
Extended Run Time (option)
Calibration:Internal calibration in approximately 10 seconds
External calibration in particle mode
Computer:Pentium processor with flash disk
Data Display:LCD, link to laptop computer for SelectScan, plasmagram, intensity map, and 3-D
view display (RS-232)
Detection Modes:Explosives - negative ion mode; Narcotics - positive ion mode