Anggota Gratis
PT HAS ENVIRONMENTAL
http://indotrade.com/pt_has_environmental/
Depan
Info Perusahaan
Katalog Produk
Hubungi Kami
Kenalkan Kami
PT HAS ENVIRONMENTAL...WE HERE TO SERVE YOU..
Katalog Produk (115)
belum tersedia
Food Contamination Kit
CHOLINESTERASE TEST KIT
Sanitarian Field Kit
Atomic Fluorescence....
Spectrophotometer AFS-....
Electronic Precision....
Soil Test Kit Cat.No : ....
Mobile Water Quality
Oil Content Analyzer
Lihat semuanya
Informasi Kontak
Nama:
Tn. Subchan Eko [Administrasi]
E-mail:
Situs Web:
www.has-environmental.com
Nomor Telpon:
Nomor Ponsel:
Nomor Faks:
Alamat:
Ruko Mega Grosir Cempaka Mas Blok i 12
Jakarta 10640, Jakarta
Indonesia
Rata-rata Tinjauan Pemakai
Tidak ada ulasan untuk perusahaan ini
Menulis tinjauan
Bahasa :
Katalog Produk
Gambar lebih besar
Dual-Ball Rotameters
Jumlah Pesanan:
Cara Pembayaran
Transfer Bank (T/T)
Negara Asal
Amerika Serikat
Kenalkan ke teman Anda
Keterangan
Rotameters are variable area flowmeters that provide field accuracy in an easy-to-read instrument. The principle of operation is simple: air flow passes through a vertical, tapered tube and pushes a ball or float having a diameter slightly less than the smaller tube end. As the ball rises, the clearance between the ball and the tube wall increases. The ball becomes stationary when the diameter of the tube is large enough to allow the total air flow past the ball. The flow rate is determined by noting the engraved scale on the tube at the position of the stabilized ball.
An important feature of the SKC dual ball rotameter is the use of two float balls: a light (black ball) for indicating low flows and a heavy (white ball) for indicating high flows. The black ball is read until it is off scale, then the white ball is read.
Two float balls give the broadest flow ranges with the highest accuracy possible in a single rotameter, The two balls give an effective reading scale length of approximately 12 inches in a shorter rotameter.
Note: Low flow rotameters can only be used with pulsation-free pumps.
Dual Ball Rotameter Operating Instructions (PDF)
About Rotameters
SKC Rotameters are secondary standards and should be calibrated by the user with a primary standard calibrator at regular intervals. Since rotameters are affected by temperature and atmospheric pressure, rotameter calibration should be performed under the conditions of use or a mathematical correction should be used.