    Datalogger Relative Humidity - Temperature - CO - CO2, Model : HD 37AB17D, HD 37B17D, Brand : DeltaOhm

    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    HD37AB17D and HD37B17D are data loggers able to measure and store at the same time the following parameters:
    • RH relative Humidity
    • T Room Temperature
    • CO Carbon Monoxide ( only HD37AB17D)
    • CO2 Carbon Dioxide

    Technical Features
    Dimensions 275 mm x 45 mm x 40 mm
    Weight 230 g ( batteries included)
    Materials ABS
    Mains power supply ( code SWD06) Batteries charger 100-240Vac/ 6Vdc-1A
    Batteries Package with 2 rechargeable batteries 1.2V
    type AA ( NiMH)
    Autonomy 8 hours of continuous working in measurement mode
    Current absorbed with 200¼ A
    instrument off
    Instrument working temperature 5° C ... 50° C
    Working relative humidity 5% RH ... 95% RH no condensed
    Temperature / Storage humidity -25° C ... + 70° C / 10% RH ... 90% RH
    no condensed
    Safety of the stored data Unlimited
    Connections
    USB interface USB 2.0 cable B type Baudrate 460800
    Charger Batteries power supply 2 - poles connector ( positive at the centre)
    ( code SWD06) Output voltage: 6Vdc
    Maximum current: 1600mA ( 9, 60 VA Max) .
    Measuring rate 1 sample every three seconds
    Storage capacity 20000 Records
    Every records includes the following:
    - date and time
    - measurement of the carbon dioxide ( CO2)
    - measurement of the carbon monoxide
    ( CO- only HD37AB17D)
    - measurement of the relative humidity ( RH)
    - measurement of the temperature ( T)
    Air Quality - CO - CO2
    Air Quality - CO - CO2
    Logging interval selectable within: 3, 6, 12, 15, 30, 60 seconds, 2, 3, 4, 5 minutes.
    The stored values represent the average value of the samples that are stored every
    three seconds.
    Printing interval selectable within: 3, 6, 12, 15, 30, 60 seconds, 2, 3, 4, 5 minutes.
    The printed values represent the average value of the samples that are stored every
    three seconds.
    Sensor Features
    Relative Humidity RH Capacitive sensor
    Sensor protection Net filter made of stainless steel ( upon request filter P6 in AISI316 sintered 20¼ m or filter P7
    in PTFE sintered 10¼ m)
    Measurement range 5...100 % RH
    Sensor working range -20...+ 60° C
    Accuracy ± 2% ( 10÷ 90% RH) ± 2, 5% in the remaining range
    Resolution 0, 1%
    Thermal effects ± 2% on all the temperature range
    Hysteresis and repeatability 1% RH
    Temperature dependence ± 2% on all temperature range
    Response time ( T90) < 20 sec. ( air speed = 2m/ sec) without filter
    Long term stability 1% / year
    Temperature T
    Sensor type NTC 10K©
    Measurement range -20...+ 60° C
    Accuracy ± 0.2° C ± 0.15% of the measure
    Resolution 0, 1° C
    Response time ( T90) < 30 sec. ( air speed = 2m/ sec)
    Long term stability 0.1° C/ year
    Carbon monoxide CO ( only HD37AB17D)
    Sensor Electro chemical cell
    Measurement range 0… 500ppm
    Sensor working range -5… 50° C
    Accuracy ± 3ppm+ 3% of the measure value
    Resolution 1ppm
    Response time ( T90) 5 years in normal environmental conditions
    Carbon dioxide CO2
    Sensor NDIR with a double wave length
    Measurement range 0… 5000 ppm
    Sensor working range -5… 50° C
    Accuracy ± 50ppm+ 3% of the measurement
    Resolution 1ppm
    Thermal effects 0, 1% f.s./ ° C
    Response time ( T90) < 120 sec. ( air speed = 2m/ sec)