    RTD – TC THERMOMETERS Pt100 AND TC INPUT, Model : HD 2178.1 HD 2178.2, Brand : DeltaOhm

    Jumlah Pesanan
    Tambah ke Keranjang
    Negara Asal
    Italia
    Kenalkan ke teman Anda Masukan ke Produk Rekanan

    Keterangan

    The HD2178.1 and HD2178.2 are portable instruments with a large LCD display. They measure the temperature using immersion, penetration air or contact probes with RTD or thermocouple sensor. A 3 or 4 wire Pt100 sensor or a 2 wire
    Pt100 sensor can be connected to the input B, while a thermocouple of type K,
    J, T, E, N to the input A.
    The probes for input B, 8 poli DIN45326 connector, are equipped with automatic
    detection module with the factory calibration settings already being memorized
    inside. A miniature polarized connector for thermocouple is provided for input A. The instrument HD2178.2 is a datalogger which stores up to 80, 000 samples. These data can be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0. The storing interval, printing, and
    baud rate can be configured using the menu.
    The models HD2178.1 and HD2178.2 are fitted with an RS232C serial port and
    can transfer the acquired measurements in real time to a PC or to a portable printer.
    The Max, Min and Avg function calculate the maximum, minimum or average
    values. Other functions include: the relative measurement REL, the HOLD function, and the automatic turning off that can also be disabled. The instruments have IP67 protection degree.

    INSTRUMENT TECHNICAL CHARACTERISTICS
    Instrument
    Dimensions
    ( Length x Width x Height) 185x90x40mm
    Weight 470g ( complete with batteries)
    Materials ABS, rubber
    Display 2x41D2 digits plus symbols
    Visible area: 52x42mm
    Operating conditions
    Operating temperature -5… 50° C
    Warehouse temperature -25… 65° C
    Working relative humidity 0… 90% RH without condensation
    Protection degree IP67
    Power
    Batteries 4 1.5V type AA batteries
    Autonomy 200 hours with 1800mAh alkaline batteries
    Power absorbed with instrument off 20¼ A
    Mains Output mains adapter 9Vdc / 250mA
    Measuring unit ° C - ° F
    Security of memorized data Unlimited, independent of battery charge
    conditions
    Time
    Date and time Schedule in real time
    Accuracy 1min/ month max departure
    Measured values storage – model HD2178.2
    Type 2000 pages containing 40 samples each
    Quantity Total of 80, 000 samples
    Storage interval 1s… 3600s ( 1hour)
    Serial interface RS232C
    Type RS232C electrically isolated
    Baud rate Can be set from 1200 to 38400 baud
    Data bit 8
    Parity None
    Stop bit 1
    Flow Control Xon/ Xoff
    Serial cable length Max 15m
    Immediate print interval 1s… 3600s ( 1hour)
    USB interface - model HD2178.2
    Type 1.1 - 2.0 electrically isolated
    Connections
    Input for probes 8 pole male DIN45326 connector
    Serial interface and USB 8-pole MiniDin connector
    Mains adapter 2-pole connector ( positive at centre)
    Measurement of temperature by Instrument – RTD sensors
    Pt100 measurement range -200 … + 650° C
    Pt1000 measurement range -200 … + 650° C
    Resolution 0.1° C
    Accuracy ± 0.05° C
    Drift after 1 year 0.1° C/ year
    Measurement of temperature by Instrument - Tc
    TC measurement range: K -200 … + 1370° C
    TC measurement range: J -100 … + 750° C
    TC measurement range: T -200 … + 400° C
    TC measurement range: N -200 … + 1300° C
    TC measurement range: E -200 … + 750° C
    Resolution 0.1° C
    Instrument accuracy
    Thermocouple K ± 0.1° C up to 600° C
    ± 0.2° C over 600° C
    Thermocouple J ± 0.1° C up to 400° C
    ± 0.2° C over 400° C
    Thermocouple T ± 0.1° C
    Thermocouple N ± 0.1° C up to 600° C
    ± 0.2° C over 600° C
    Thermocouple E ± 0.1° C up to 300° C
    ± 0.2° C over 300° C
    The accuracy only refers to the instrument. Error due to the thermocouple
    or to the cold junction reference sensor is not included.
    Temperature drift @ 20° C 0.02% / ° C