BINDER incubator - BD series[30 Sep. 2011, 5:13:42]
Overview
ncubator with gravity convection
The BD series BINDER incubator with gravity convection are specially designed for long-term and stable continuous operation. Ideal for gentle incubation of organisms, as well as conditioning of heat-sensitive media or microbiological heating.
Performance Characteristics
Electronically controlled APT.lineâ„¢ preheating chamber assuring temperature accuracy and reproducible results
Temperature range from 5 ° C ( 9 ° F) above room temperature to 100 ° C ( 212 ° F)
DS control with integrated timer 0 to 99 hrs
Digital temperature setting with an accuracy of a tenth of a degree
Independent temperature safety device class 3.1 ( DIN 12880) with optical alarm
Adjustable ventilation by means of front ventilation flap slide and rear exhaust Ø 50 mm
Inner glass door
RS 422 interface for communication software APT-COMâ„¢ DataControlSystem
Units up to 115 liters are stackable
2 chrome-plated racks included
BINDER test confirmation
BD 23
Exterior dimensions
Width ( mm) 435
Height ( incl. feet) ( mm) 495
Depth ( mm) 520
Plus door handle and connection ( mm) 90
Wall clearance, rear ( mm) 100
Wall clearance, side ( mm) 100
Steam space volume ( l) 36
Number of doors ( ea.) 1
Interior dimensions
Width ( mm) 222
Height ( mm) 330
Depth ( mm) 277
Interior volume ( l) 20
Racks ( number standard/ max.) 2 / 3
Load per rack ( kg) 12
Permitted total load ( kg) 25
Weight ( empty) ( kg) 26
Temperature data
Temperature range from 5 ° C above ambient temperature to ( ° C) 100
Temperature variation
at 37 ° C ( ± K) 0, 5
at 50 ° C ( ± K) 1, 8
Temperature fluctuation
at 37 ° C ( ± K) 0, 2
at 50 ° C ( ± K) 0, 3
Warm-up time 1)
to 37 ° C ( min.) 49
to 50 ° C ( min.) 51
Recovery time after doors were open for 30 sec. 1)
at 37 ° C ( min.) 3
at 50 ° C ( min.) 4
Electrical data
IP protection class acc. to EN 60529 IP 20
Voltage ( ± 10% ) 50/ 60 Hz ( V) 230
Nominal power ( kW) 0, 2
Energy consumption at 37 ° C ( W) 11
1) To 98% of the set value
All technical data are specified for units with standard equipment at an ambient temperature of 25 ° C and a line voltage fluctuation of ± 10% . The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.