KYORITSU 2413F[16 Mar. 2009, 2:15:34]
Keterangan
� Extra wide transformer jaws are best suited for clamping on all three or four wires (3phases) together for leakage current measurement.
� Frequency filter switch to eliminate the effect of harmonics.
� Peak hold function.
� Analogue output terminal.
AC A
(50/60Hz) : 200mA/2/20/200A/1000A
|±1.5%rdg±2dgt (200mA/2/20A)
|±2%rdg±2dgt (200A 0~500A)
|±5.5%rdg (501~1000A)
AC A
(WIDE) : 200mA/2/20/200/1000A
|±1%rdg±2dgt [50/60Hz],
|±3%rdg±2dgt [40Hz~1kHz] (200mA/2/20A)
|±1.5%rdg±2dgt [50/60Hz],
|±3.5%rdg±2dgt [40Hz~1kHz] (200A/0~500A)
|±5%rdg [50/60Hz],
|±10%rdg [40Hz~1kHz] (501~1000A)
Conductor Size : �68mm max.
Effect of External
Stray Magnetic Field
�15mm 100A : 10mA AC max.
Frequency Response : 40Hz~1kHz
Output : AC/DC 200mV against 2000 count
Withstand Voltage : 3700V AC for 1 minute
Applicable Standard : IEC 61010-1 CAT.III 300V
IEC 61010-2-032
Power Source : 6F22 (9V) � 1 *Continuous measuring time: Approx. 60 hours
Dimensions : 250(L) � 130(W) � 50(D)mm
Weight : 570g approx.
Accessories : 9094 (Carrying Case)
6F22 � 1
Instruction Manual
Optional : 7073 (2WAY Output Cord)