Keterangan :
WFX-110B/ 120B/ 130B Flame Atomic Absorption Spectrophotometer ( AAS )
Manufacturer : RAYLEIGH
ISO 9001: 2000 & ISO 4001: 2004
CE-TUV GmbH
Features:
1 Innovated rich oxygen air-acetylene flame analysis technique ( WFX-110B ) :
The patented flame analysis technique adopting rich oxygen air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses, such as Ca, Al, Ba, W, Mo, Ti, V, etc. Flame temperature is continuously adjustable between 2300-2950¡ æ , with makes it possible to choose the best atomization temperature for different elements. It features easy operation, low analysis cost and wide flame AAS analytical range. Rich oxygen flame will not pollute the environment and is not harmful to human bodies. It' s break-through in flame AAS analysis
2 Integrated flame/ graphite furnace atomization system, changeable with flame emission burner
2.1 Automatically controlled changeover of the integrated flame and graphite furnace atomizer featuring easy operation and time saving eliminates human labor
2.2 A flame emission burner head can be installed to perform flame emission analysis to alkali metals as K, Na etc. ( WFX-110B/ 120B )
3 Accurate fully automated control system
3.1 Automatic multi-lamp turret, automatic adjustment of lamp current and optimization of light beam position
3.2 Automatic wavelength scanning and peak picking
3.3 Automatic spectral bandwidth changing
3.4 Automatic ignition
4 Perfect safety protection measures
4.1 Alarm and automatic protection to fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
5 Advance and reliable electronic design
5.1 Adopting large-scale programmable logic array and Inter I2C bus technology
5.2 European type sockets and AMP adapters with high reliability to ensure long term reliability of the whole electronic system
6 Easy and practical analysis software
6.1 Easy-to-use AAS analysis software is made under Windows operating system, realizing fast parameter setting and optimization
6.2 Automatic sample dilution, automatic curve fitting, automatic sensitivity correction
Main Specification
Wavelength range 190-900nm
Wavelength accuracy Better than ¡ à 0.25nm
Resolution Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30% .
Baseline stability 0.004A/ 30min
Background correction The D2 lamp background correction capability at 1A is better than 30 times
The S-H background correction capability at 1.8A is better than 30 times.( only for WFX-110B/ 120B
Light Source
Lamp turret 6-lamp turret ( WFX-110B/ 120B )
Lamp turret 4-lamp turret ( WFX-130B) ; Auto-alignment, fully automated scan and peak-picking.
Lamp current adjustment Automatic Adjustment and display
Wide pulse current: 0~ 25mA, Narrow pulse current: 0~ 10mA
Lamp power supply mode 400Hz square wave puls
100Hz narrow square wave
pulse+ 400Hz wide square wave pulse ( WFX- 110B/ 120B)
Optical System
Monochomator Single beam, Czerny-Turner design grating monochromator
Grating 1800 l/ mm
Focal length 277mm
Blazed Wavelength 250nm
Spectral Bandwidth 0.1nm, 0.2nm, 0.4nm, 1.2nm, automatic change
Flame Atomizer
Burner 10cm single slot all-titanium burner
Spray chamber Corrosion resistant all-plastic spray chamber
Nebulizer High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7ml/ min
Emission burner provided with WFX-110A/ 120A
Detection and Data Processing System
Detector R928 Photomultiplier with high sensitivity and wide spectral range
Software Windows operating system
Analytical method Working curve auto-fitting; standard addition method; automatic sensitivity correction, automatic calculation of concentration and content.
Repeat times Maximum 20 times of repeat measurement, automatic calculation of mean value, standdard deviation and relative standard deviation.
Multi-task Functions Sequential measurement for multi-element determination in one sample
Condition reading With model function
Result printing Measurement data and final analytical report printout, editing with Excel
Standard RS-232 serial port communication
Characteristic Concentration and Detection Limit
Normal Air-C2H2 flame Cu: Characteristic concentration¡ à 0.025mg/ L, Detection limit¡ à 0.006mg/ L;
Oxygen-rich Air-C2H2 flame Ba: Characteristic concentration ¡ à 0.22mg/ L, Al: Characteristic concentration ¡ à 0.4mg/ L ( for WFX-110A )
Function Expansion Hydride vapor generator can be connected for hydride analysis
Dimensions and weight
Main unit 102¡ à 49¡ à 54cm, unpacked 80kg