AA-3500 Air-Aide Airborne Particulate Monitor[21 May. 2011, 6:17:46]
Air-Aide Airborne Particulate Monitor Model AA-3500 is a portable handheld survey tool for measuring lung damaging particulate matter. The unique sensor design has applications in indoor air quality, HVAC, environmental and industrial hygiene.
Applications
AA-35000 can be used for all sorts of particulate monitoring. The AA-3500 has several application uses seen below that can be helpful in developing an air regulation system. It can be used for both indoor ambient monitoring. The AA-3500 is an excellent assistant tool for the OSHA/ EPA/ ASHREA compliance methods. Giving you direct real-time readings in mg/ m³ .
* Lead
* Welding Fumes
* Silica
* Coal dust
* Wood/ Paper
* Concrete/ Cement
* Cadmium
* Nuisance Dusts
* Chromate
* Toxic Soil Remediation
* Tobacco smoke
* Construction Dust
* Dry Chemical
* Oil Mists
* Grinding Dusts
* Pharmaceuticals
* Spray Paint
* Grain
Features
The Air Aide Airborne particulate monitor has a multitude of features the list could go on forever, but for now we will just name a few. The AA-3500 immediately displays and stores data into the memory. The supplied PC software allows for graphical and statistical representation of the total dust generating process, eliminating the need for repetitive and costly filter gravimetric analysis.
* Direct reading with high resolution and stability.
* Active sample flow using a long-life air sampling pump.
* Location codes; user-defined audible alarm; security code restricts unauthorized use.
* Size Selective Impactors
* Rechargeable battery. Units run on AC or DC power.
* On-screen programming of sampling and data storage parameters.
* Selection of different built in-calibration factors for fine and coarse particulate matter.
* CE, ETL approved.
Specifications
Display: Alpha-numeric LCD, 4 line 20 character mg/ m³ concentration reading
Operation: Four key splash proof membrane switch
Calibration: NIOSH gravimetric method - NIST traceable - SAE fine test dust
Accuracy: + / - 10% to filter gravimetric SAE fine test dust
Sensing range: 0.001 to 20.0 mg/ m³ or .01 to 200 mg/ m³ ( optional)
Particulate size range: .01 - 100 µ m
Precision: ± .003 mg/ m³ ( 3 µ g/ m³ )
Sampling flow rate: 2 Lpm
Analog output: 0-2 vdc
Alarm output: 90dB at 3 ft
Record time: 1 second to 15 months
Sampling rate: Short Term - 2, 4, and 10 seconds
Long Term - 1, 10 and 30 minutes
Data storage: 21, 600 data points
Memory & time storage: > 5 years
Real-time clock & display: hours, minutes, seconds, day, month, year
Data display: concentration in mg/ m³ & TWA, Maximum, Minimum, STEL, date, time
Digital output: RS-232
Operation temperature: -10° C to 50° C
Storage temperature: -20° C to 70° C
DustComm Pro Software: Windows driven for graphical and data translation
Power: Rechargeable battery
Operating time: > 10 hours on fully charged battery
Charging time: